Abbemat Essential 3001

  • Accuracy: ±0.0001 nD  / ± 0.05 °Brix
  • Measuring range: 1.33 nD to 1.55 nD / 0 °Brix to 100 °Brix
  • Precise temperature control: 20 °C to 40 °C
  • 10 pre-installed measuring scales
  • Print results with optional label printer
For Additional Information Request a Quote
Category:

Description

Sign your results with confidence: The Abbemat Essential line is a budget-friendly benchtop refractometer providing best-build quality, precision, and temperature control. Measure your sample with just one click, print out and sign the results, and directly insert into your laboratory book – it’s that simple and leaves no room for transcription errors. All three automatic benchtop refractometers are suitable for liquids, viscous or solid substances, and turbid solutions. No training is needed; efficient use by all staff members is possible.

Key features

Proven Peltier temperature control for all of our benchtop refractometers

Precise temperature measurement and control are prerequisites for accurate refractive index measurements. The patented T-Check system that is used for each device before it is delivered ensures surface temperature measurement and control accuracy of your automatic benchtop refractometer of up to ±0.03 °C.

  • Traceable temperature check and adjustment
  • Most precise prism-surface temperature measurement
  • Prevention of temperature-induced errors

Fast concentration measurements

Speeding up of concentration measurements of binary solutions due to a simplified user interface and predefined methods.

  • Refractive index and Brix measurements
  • Eight additional methods (exchangeable)
  • Short cycle times down to 10 seconds to measure and document results
  • Low sample volume of 70 µL

Scratch-proof sapphire prism for long-term precise results

The sapphire prism each benchtop refractometer is equipped with guarantees the most precise results throughout the entire instrument lifetime.

  • Resists cleaning with spatula or other lab tools
  • Mohs hardness of 9, close to diamond-grade
  • Backed by a 25-year warranty on the prism

Easy printing and documentation of results

The Abbemat Essential benchtop refractometers simplify results documentation with their optional label printer, allowing for fast and reliable printouts. This eliminates the need for manual data copying, reducing the risk of transcription errors, and ensuring that all recorded measurements are accurate and traceable.

Ready-to-use right out of the box

Our automatic benchtop refractometers are designed for immediate use with a no-frills user interface that requires no setup. The Abbemat Essential’s intuitive operation ensures that even first-time users can perform measurements effortlessly without the need for training. This minimizes errors and maximizes efficiency in any laboratory environment.

Specifications

Abbemat 3001 Abbemat 3101 Abbemat 3201
Measuring specifications
Refractive index
Range 1.33 nD to 1.55 nD 1.30 nD to 1.66 nD 1.26 nD to 1.72 nD
Max. resolution 0.0001 nD 0.00001 nD 0.000001 nD
Accuracy1) ±0.0001 nD ±0.00004 nD ±0.00002 nD
Repeatability ±0.0001 nD ±0.0000015 nD ±0.0000015 nD
Brix
Range 0 °Brix to 100 °Brix 0 °Brix to 100 °Brix 0 °Brix to 100 °Brix
Display resolution 0.01 °Brix 0.001 °Brix 0.001 °Brix
Accuracy ±0.05 °Brix ±0.03 °Brix ±0.015 °Brix
Temperature
Temperature range 20 °C to 40 °C 10 °C to 70 °C 42) °C to 105 °C
Temperature probe accuracy1) ±0.05 °C ±0.03 °C ±0.03 °C
Temperature probe stability ±0.002 °C ±0.002 °C ±0.002 °C
Display 7″ Hybrid touch
Communication interfaces USB (printer)
Data storage Label printer
General
Min. sample volume 70 uL
Max. allowed pressure in micro flow cell 10 bar
Ambient temperature 5 °C to 40 °C
Materials in contact with the sample
Prism Synthetic sapphire
Sample well Stainless steel, optionally Ni-alloy
Seal FFKM (perfluoroelastomer)
Compounds
Light source LED light source, average lifetime > 100,000 hours
Wavelength 589.3 nm (by wavelength-adjusted interference filter)
Power requirements 100 VAC to 240 VAC +10 %/-15 %, 50/60 Hz, min. 10 W, max. 100 W,
depending on sample temperature setting and ambient temperature
Dimensions
W x H x D 225 mm x 115 mm x 330 mm
Weight 3.5 kg

1) Valid at standard conditions (T=20 °C, ambient temperature 20 °C to 25 °C
2) At max. ambient temperature of 35 °C