A steel test cup rotating at 800 RPM is pressed against a steel test block. Sample under test is carried by the test cup into the sliding contact. Test load at the contact is progressively increased, score value and OK value are determined.
"Compliant to ASTM D2509 and D2782 Fluid recirculation system Automatic grease feeder Friction lever and weights 1:10 Loading lever with load cell and digital indicator Oil reservoir with heating system"
Rotational Speed: 100 to 800 RPM Rate of Loading: 0.9 to 1.3 kg/sec Dead Weights: 52 kg Auto Fluid Recirculation Feeder: 3L capacity Auto Grease Feeder: 45 ± 10 g/min Preset Timer: 99hrs:59min:59sec Preset Temperature Controller: up to 100°C Loading Arm: 1:10 Ratio